Organic X Monitor HC

Continuous detection of volatile organic compounds
Environmental pollution
with volatile organic compounds (VOC, POC) must be avoided and contaminated
sites have to be cleaned up. This makes a robust, reliable and sensitive monitor
to detect this compounds most desirable.
On application of
the Organic X Monitor HC numerous organic compounds can be easily
detected as summary parameter:
· Halogenated Hydrocarbons
like CFC, Vinyl Chloride, Chlorobenzene etc.
· BTEX
· Fuels
· Styrene
· Naphthalene
· Alcohols
· Many others
The Organic X Monitor HC developed and patented by Degussa-Hüls AG meets all requirements of a successful continuous monitoring device:
· High Sensitivity
· Short Response
Time
· Easy to Handle
· Robustness
· Low Maintenance
Operational Principle
Volatile organic compounds are stripped off the liquid matrix by a stream of clean gas and their inorganic products are detected by measuring the conductivity after UV-oxidation.
Fields of Application
The Organic X Monitor HC can be applied for the detection of volatile organic compounds in a wide range of applications like:
· Potable Water Production
· Wastewater Monitoring
· Cooling Water Monitoring
· Discharge Monitoring
· Soil- and Groundwater
Remediation
· Carbon Break-through
It should be emphasized that even in highly contaminated waste water in most cases no sample pretreatment is necessary.
High Sensitivity
The Organic X Monitor HC is equipped with high sensititve conductivity cells. Variations in temperature are compensated by a polynome based temperature compensation. This results in a lower detection limit of less than 0.5ppb! Dependend on the parameter setting analytes can be measured in a wide range.
Low Maintenance
Less than 20h annually are required for maintenance. The availability is almost 99.8%! Thanks to the modular design maintenance is as easy as anything and can be performed by the user. Each component is readily accessible and in most cases no tools are required for replacements. In order to maximize safety the electronic compartment is separated from the wet-chemistry part of the monitor.
Robustness
All components are chosen for a maximum of robustness to withstand harsh environmental conditions typical for process applications.
Modular Design
The rack mount modular design makes it very easy to customize the monitor to special monitoring tasks. Also other moduls like process PC´s, chart and paperless recorders can be incorporated.
Specification Organic X Monitor HC
Sample: Fluid samples
Sample flow rate:
0.5 - 10l/h, not pressurized
Sample temperature:
< 30°C, higher temperatures with optional sample cooler
Range: Conductivity
after UV oxidation: 0 – 2µ
S/cm or 0 - 20µ
S/cm. The actual range of the analyte concentration may vary in wide extents.
Detection Limits:
0,5µ
g/l Perchlorethylene, 10µ
g/l Diesel, 20µ
g/l Naphthalene
Power Requirements:
230V +/-10%; 50Hz; 260 VA
Outputs: 4
isolated analog outputs 4-20 mA (0 – 20mA), for conductivity and temperature.
Max. load 600W .
Communication:
RS 485 and Bus operation optional
Alarm: Alarm
relay; thresholds user adjustable
Display: 2
LCD multi-displays (each for conductivity and temperature)
Case: 44 x
60 x 101 cm (DxWxH), steel case
Weight: 80kg
Degree of Protection:
IP 64
Ambient Conditions:
Temp.: +5 - +40°C; rel. humidity: <95% not condensing